High Voltage / Piezo Amplifiers & Signal Conditioners
Fiber Optic Links / Data Acquisition Systems
RF Power Amplifiers / Dual Directional Couplers
Instruments for Biological Science
Under Water Metal Detectors, Camera’s & Sonar Systems
|
Gas Analyzers for Animals / Insects / Plants / Human / Environmental:
S128 Methane Analyzer 5000ppm
Qubit's S128 Methane Analyzer uses state of the art Non-Dispersive Infrared Technology (NDIR) to measure Methane Levels in a flowing gas at two ranges 0-1250ppm and 0-5000ppm. The instrument is equipped with a Digital Display that provides a measure of the methane concentration in units of ppm. In addition, an Analog Output Signal (0-5VDC) is provided for Data Logging by a computer with our Optional Software and Data Interface.
The S128 Methane Analyzer is a Flow through Gas Analyzer therefore gas should be supplied under pressure at flow rates not to exceed 650ml/min. An optional built-in Gas Pump may be provided with the S128-P unit for easy delivery of the gas to the sensor. Qubit' S128 Methane Analyzer may be used in the field with Optional Battery Pack (A249).
- Features:
- Flow-through Infrared Gas Analyzer
- Dual Range Operation: 0-1250ppm and 0-5000ppm
- Easy Two Point Calibration
- Digital Display
- 0-5V Analog Output
- Optional Built-in Gas Pump (S128-P)
- Applications:
- Fermentation Studies
- Waste Management
- Bioreactors and Biofuels Research
- Single Animal Gas Exchange
- Livestock Emissions
- Land Emissions
- Specifications:
- Operating Principle: Non-Dispersive Infrared
- Gas Sampling Mode: Flowing Gas Stream, Sealed Chamber
- Maximum Gas Flow Rate: 650mL/min
- Measurement Range: (LCD display) 0-5000ppm
- Analog Output, Low Sensitivity: 0-5000ppm
- Analog Output, High Sensitivity: 0-1250ppm
- Accuracy (assumes accurate recent calibration): ±40ppm CH4
- Repeatability (assumes stable atm press and temp): ±40ppm CH4
- Maximum Drift (per year): ±100ppm (STP)
- Response time (@250 mL/min; to 95% of final value): 45sec
- Warm up time (@22°C): 5 min
- Output (Linear) for Low Sensitivity Setting: 0-5 VDC for 0-5000ppm
- Output (linear) for High Sensitivity Setting: 0-5 VDC for 0-1250ppm
- Calibration Adjustments: Zero and Span
- Two Point Calibration
- Operating Temperature Range: 0 to 45°C
- Storage Temperature Range: -40 to 50°C
- Operating Pressure Range: ±1.5% Local Mean Pressure
- Humidity Range: 5 to 95% RH, Non-Condensing (recommend drying gas stream)
- Pressure Dependence: +0.19% reading per mm Hg
- Power Requirements: 12 VDC via 100/240 VAC 60Hz Adapter
- Current Requirements: 175mA Average, 450mA Peak
- Dimensions (H × W × D) (cm): 12.8 × 28.2 × 25.6
- Weight: 3.0 kg
- Warranty: 1 Year Limited
All Specifications are subject to change without a Prior Notice
For any sort of clarifications kindly contact: info@globesolutionz.com
|