Gas Analyzers for Animals / Insects / Plants / Human / Environmental:


S128 Methane Analyzer 5000ppm


Qubit's S128 Methane Analyzer uses state of the art Non-Dispersive Infrared Technology (NDIR) to measure Methane Levels in a flowing gas at two ranges 0-1250ppm and 0-5000ppm. The instrument is equipped with a Digital Display that provides a measure of the methane concentration in units of ppm. In addition, an Analog Output Signal (0-5VDC) is provided for Data Logging by a computer with our Optional Software and Data Interface.

The S128 Methane Analyzer is a Flow through Gas Analyzer therefore gas should be supplied under pressure at flow rates not to exceed 650ml/min. An optional built-in Gas Pump may be provided with the S128-P unit for easy delivery of the gas to the sensor. Qubit' S128 Methane Analyzer may be used in the field with Optional Battery Pack (A249).

  1. Features:
  2. Flow-through Infrared Gas Analyzer
  3. Dual Range Operation: 0-1250ppm and 0-5000ppm
  4. Easy Two Point Calibration
  5. Digital Display
  6. 0-5V Analog Output
  7. Optional Built-in Gas Pump (S128-P)
  1. Applications:
  2. Fermentation Studies
  3. Waste Management
  4. Bioreactors and Biofuels Research
  5. Single Animal Gas Exchange
  6. Livestock Emissions
  7. Land Emissions
  1. Specifications:
  2. Operating Principle: Non-Dispersive Infrared
  3. Gas Sampling Mode: Flowing Gas Stream, Sealed Chamber
  4. Maximum Gas Flow Rate: 650mL/min
  5. Measurement Range: (LCD display) 0-5000ppm
  6. Analog Output, Low Sensitivity: 0-5000ppm
  7. Analog Output, High Sensitivity: 0-1250ppm
  8. Accuracy (assumes accurate recent calibration): ±40ppm CH4
  9. Repeatability (assumes stable atm press and temp): ±40ppm CH4
  10. Maximum Drift (per year): ±100ppm (STP)
  11. Response time (@250 mL/min; to 95% of final value): 45sec
  12. Warm up time (@22°C): 5 min
  13. Output (Linear) for Low Sensitivity Setting: 0-5 VDC for 0-5000ppm
  14. Output (linear) for High Sensitivity Setting: 0-5 VDC for 0-1250ppm
  15. Calibration Adjustments: Zero and Span
  16. Two Point Calibration
  17. Operating Temperature Range: 0 to 45°C
  18. Storage Temperature Range: -40 to 50°C
  19. Operating Pressure Range: ±1.5% Local Mean Pressure
  20. Humidity Range: 5 to 95% RH, Non-Condensing (recommend drying gas stream)
  21. Pressure Dependence: +0.19% reading per mm Hg
  22. Power Requirements: 12 VDC via 100/240 VAC 60Hz Adapter
  23. Current Requirements: 175mA Average, 450mA Peak
  24. Dimensions (H × W × D) (cm): 12.8 × 28.2 × 25.6
  25. Weight: 3.0 kg
  26. Warranty: 1 Year Limited

All Specifications are subject to change without a Prior Notice

For any sort of clarifications kindly contact:
info@globesolutionz.com